Classic Scientific make Photo Stability Chamber is Designed to suit standards for photogenic effect along with temperature & humidity for not less than 1.2 million Lux hours and ultra violet source of energy for not less Than 200 watts hours / sq. Meter. As per ICH & C gmp Guidelines. Range:- Temperature range : 10 Deg C to 50 Deg C. Temperature accuracy : +/- 0.2 Deg C. Temperature Uniformity : +/- 1 Deg C. Humidity range : 40% R.H. to 80 % R.H. Humidity accuracy : +/- 2% R.H. Humidity uniformity : +/- 3% R.H. Application : PHOTO STABILITY CHAMBER are used for studies related to photogenic effect on the samples exposed to U.V & day light illuminations. It is design to create natural climatic conditions with the help of dry air, refrigeration & Humidity system built-in this chamber are basically in application with pharmaceuticals & also electro-mechanical R & D test to confirm the successful result of the product under test.